DISPLAY RESULT OF “P” BUTTON
“Pro” <>“0” - Causes the indicator to return to normal display mode.
Any changes to set-up data are permanently stored in the
E
2
PROM.
“Pro” <>“1” - Entry into this module allows the user to select the
Thermistor type, whether the display will read in degrees
Fahrenheit (F) or Celsius (C), and display decimal point
position.
“Pro” <>“2” - Entry into this module allows the user to select
non-standard display slope and display offset values.
This enables the meter to be “scaled” to a calibrated
temperature probe. (This scaling is NOT required for
most applications.)
“Pro” <>“3” - Module #3 allows the user to program what can be
accessed from the front panel when the PGM. DIS.
(Program Disable, TBA #3) pin is connected to common.
This feature protects critical set-up data from accidental
modification while allowing access to setpoints and other
functions. The front panel lock-out menu (quick
programming) includes setpoint modification,
integrator/totalizer resetting, and peak/valley resetting.
Note: The term “Quick Programming” is used to refer to the
ability to change the information that can be accessed from
the front panel when the “PGM. DIS.” terminal is connected
to “COMM.”.
“Pro” <> “4” - Module #4 programs the digital filtering level and the
function of the remote input “E1-CON” pin (TBA #4), and
if the totalizer option is installed, the remote input
“E2-CON” pin (TBA #8). The functions of the remote E1
and E2 pins are the same and include display hold,
peak/valley modes, totalizer reset, alarm reset, temperature
offset, reading synchronization or print request.
DISPLAY
RESULT OF “P” BUTTON
“Pro” <>“5” - This module sets the time base, scale factor and low
temperature disable function for the optional integrator/
totalizer.
“Pro” <>“6” - This module allows programming for the basic
configuration of the alarm option. The programming
includes HI/LO acting, tracking, alarm display, latched
or auto-reset, assignment to either the input or the
integrator/totalizer, and alarm and hysteresis values.
“Pro” <>“7” - Module #7 is the serial communication parameter
programming. Baud rate, unit address, print request
function and condensed prints are all programmable.
“Pro” <>“8” - This module allows digital scaling of the retransmitted
analog output. Display values that correspond to 4 mA or
0 VDC and 20 mA or 10 VDC are keyed-in to scale the
output and it may be assigned to either the input or the
integrator / totalizer.
“Pro” <>“9” - This module is the service operation sequence and is not
normally accessed by the user. This step re-calibrates the
basic input and is used to compensate for long-term drift.
Execution of this module should be done by technicians
with the proper equipment in accordance with a
maintenance plan of yearly recalibrations. A code number
entry step is used to protect from inadvertent entries. Also,
there is a number of other access codes which provide test
and set-up changes as an aid in troubleshooting.